A pH meter measures how acidic or alkaline a water-based sample is, on a scale that most commonly runs from 0 to 14, using a glass electrode, a reference and a temperature sensor. Choosing one comes down to four things: form factor (pocket tester, portable meter or bench meter), accuracy and resolution, an electrode that suits your sample, and how much calibration and care you are willing to carry. This guide compares the pH instruments ERE carries from Hanna Instruments and Ohaus, using the calibration and electrode rules published by USGS, CCME and Health Canada.
In this guide
- What a pH meter measures
- Canadian pH guideline levels
- Pocket tester, portable meter or bench meter
- The pH meters ERE carries, compared
- Which pH meter is the most accurate
- Choosing the pH electrode
- Calibrating a pH meter
- Taking a reliable pH reading
- Electrode care and storage
- Where pH fits in field and wastewater work
What Does a pH Meter Measure?
A pH meter measures the hydrogen-ion activity of a water-based sample and displays it as a pH value. CCME's pH fact sheet defines pH as the negative base-10 logarithm of the hydrogen-ion activity: 7.0 is neutral, lower values are acidic, higher values are alkaline, and each unit is a tenfold change in hydrogen-ion concentration.
Inside the meter, the glass electrode develops a small voltage that depends on the sample, and the meter converts it to pH using its last calibration. The USGS National Field Manual chapter on pH gives the theoretical response as 59.16 mV per pH unit at 25 °C, and notes that the response changes with temperature, which is why every instrument below has temperature compensation. Temperature also shifts pH itself: USGS calculates neutral pure water at 6.92 at 30 °C and 7.48 at 0 °C, so a pH value should always be recorded with the sample temperature.
What pH Levels Do Canadian Guidelines Refer To?
The three reference points most Canadian field programs work against are CCME's freshwater and marine aquatic-life guidelines and Health Canada's drinking-water range.
| Reference | Value | What it covers |
|---|---|---|
| CCME, freshwater | 6.5 to 9.0 | Long-term guideline for the protection of freshwater aquatic life |
| CCME, marine and estuarine | 7.0 to 8.7 | Unless the pH is naturally outside this range; within it, pH should not vary by more than 0.2 units from the natural pH expected at that time |
| Health Canada, drinking water | 7.0 to 10.5 | Operational guideline for finished drinking water, set to support treatment effectiveness and corrosion control |
Permit conditions and provincial criteria for your site may differ, so confirm the limit that applies before you choose an instrument. It also sets the accuracy you need: checking against a 6.5 to 9.0 range does not call for the same resolution as work that tracks changes of a few hundredths of a pH unit.
Pocket Tester, Portable Meter or Bench Meter: Which Should You Choose?
Choose a pocket tester for quick spot checks, a portable meter with a cable probe for field work and logging, and a bench meter when the sample comes to the lab. ERE carries all three from Hanna Instruments and Ohaus.
Pocket testers
A pocket tester is a pen-style meter with the electrode built in. The pHep 4 and pHep 5 use a replaceable pull-out cartridge electrode with the temperature sensor exposed next to the glass for fast, accurate compensation, according to the ERE product record. The Hanna Combo HA-98129 and HA-98130 add electrical conductivity (EC), total dissolved solids (TDS) and temperature in one floating, waterproof unit with a replaceable graphite probe.
Portable and multiparameter meters
Portable meters put the electrode on a cable, add a screen you can read in the field and many log data. The Hanna HI991300 and HI991301 read pH, EC, TDS and temperature through one probe on a 1 m cable and ship in a hard carrying case. The Hanna edge is a 12 mm slim meter with digital electrodes that measures pH, conductivity or dissolved oxygen, with data logging and five-point calibration. The Hanna HI9829 puts pH and oxidation-reduction potential (ORP) alongside EC, dissolved oxygen and turbidity on a submersible probe. For soil, the Hanna HA-99121 kit includes an auger so pH can be read directly in the ground, and sample preparation solutions are available for diluted soil samples.
Bench meters
Bench meters trade portability for a stable electrode holder and, in the higher-end models, finer resolution. The Ohaus Aquasearcher AB23PH is the simple pH and ORP model, the AB33PH adds a touch keypad and a wider range, and the AB41PH is the research-grade model at 0.001 pH resolution. Each Ohaus bench meter is sold either as the meter alone (the -B part number) or with an electrode and buffer mini kits (the -F part number), so ask which configuration a quote includes.
How Do the pH Meters ERE Carries Compare?
The table sets the instruments side by side. Every value comes from the manufacturer's data in ERE's product record.
| Instrument | Type | pH range | Also worth knowing |
|---|---|---|---|
| Hanna pHep 4 (HA-98127) | Pocket tester | −2.0 to 16.0 | ±0.1 pH, 0.1 pH resolution; replaceable pull-out cartridge electrode; automatic 1- or 2-point calibration |
| Hanna pHep 5 (HA-98128) | Pocket tester | −2.0 to 16.0 | ±0.05 pH; same cartridge electrode and calibration as the pHep 4 |
| Hanna Combo HA-98129 / HA-98130 | Pocket tester: pH, EC, TDS, temperature | 0.00 to 14.00 | Floating waterproof case; replaceable graphite probe; EC to 3,999 µS/cm (98129) or 20.00 mS/cm (98130) |
| Hanna HI991300 / HI991301 | Portable: pH, EC, TDS, temperature | 0.00 to 14.00 | Probe on 1 m cable; hard case; automatic 1- or 2-point pH calibration; the 991301 reads higher-range conductivity |
| Hanna edge | Portable, digital electrodes | −2.000 to 16.000 | ±0.002 pH at 0.001 resolution (±0.01 at 0.01); five-point calibration; CAL Check and Sensor Check indicators; data logging; 250 g |
| Hanna HI9829 | Portable multiparameter, submersible probe | 0.00 to 14.00 | ±0.02 pH at 0.01 resolution; ORP ±2,000 mV; IP67 meter, IP68 probe; logs up to 44,000 samples |
| Hanna HA-99121 soil kit | Portable kit for soil | −2.00 to 16.00 | ±0.02 pH at 0.01 resolution; HA-1292D electrode on 1 m cable; auger for direct soil measurement |
| Ohaus Starter 2200 | Bench | 0.00 to 14.00 | ±0.01 pH at 0.01 resolution; also mV/ORP and temperature; auto-buffer recognition and auto-endpoint; 99-item memory |
| Ohaus Aquasearcher AB23PH | Bench | 0.00 to 14.00 | ±0.01 pH at 0.01 resolution; pH and ORP; 3-point calibration with auto buffer recognition; 99-item memory |
| Ohaus Aquasearcher AB33PH | Bench | −2.00 to 16.00 | 0.01 pH resolution; touch keypad; large LCD |
| Ohaus Aquasearcher AB41PH | Bench, research-grade | −2.000 to 20.000 | 0.001 pH resolution; the -F kit adds an ST410 pH electrode, STTEMP probe and compact stirrer |
If you also need conductivity, dissolved oxygen or turbidity, our water quality meter guide covers multiparameter instruments as a class.
Which pH Meter Is the Most Accurate?
Of the meters listed here, the Hanna edge has the tightest published figure, ±0.002 pH at 0.001 resolution, followed by the Ohaus Starter 2200 and AB23PH at ±0.01 pH, the HI9829 at ±0.02 pH and the pHep 5 at ±0.05 pH. Health Canada's pH document puts the field-versus-lab split in the same terms: pocket-sized, battery-powered portable meters are "suitable for field use" at an accuracy of ±0.05 pH units, while larger, more sophisticated models can attain ±0.01 pH units.
Resolution is not accuracy. A display that shows three decimals says how finely the meter reads, not how close it is to the true value, so compare the accuracy line on the datasheet. The number also depends on the buffers. Hanna's pH 4.01, 7.01 and 10.01 calibration solutions are specified to ±0.01 pH at 25 °C and traceable to NIST standard reference materials, and Hanna's millesimal solutions reach ±0.002 pH.
Finally, match the figure to how the result is used. USGS records field pH to the nearest 0.01 unit and reports it to the nearest 0.1 unit, so a meter with 0.01 resolution and a well-run calibration already covers a standard field protocol. The finer instruments earn their place in the lab and in research work.
Which pH Electrode Does Your Sample Need?
Choose the electrode for the sample first, then the meter. Hanna's pH and ORP electrodes are combination sensors that typically come with a BNC connector and 1 m cable, and the recommended use differs by model:
| Model | Recommended use | pH range | Temperature | Reference |
|---|---|---|---|---|
| HA-1043B | Hydrocarbons, paints, solvents, sea water, strong acids and bases, high-conductivity samples | 0 to 14 | 0 to 100 °C | Double junction, Ag/AgCl |
| HA-1053B | Soil samples, potable water, emulsions, fats and creams, semi-solid products (conical bulb) | 0 to 12 | −5 to 100 °C | Single junction, Ag/AgCl |
| HA-1131B | General laboratory use | 0 to 13 | −5 to 100 °C | Single junction, Ag/AgCl |
| HA-1230B | Field applications | 0 to 13 | 0 to 80 °C | Double junction, Ag/AgCl |
Two sample types need extra care. USGS notes that most standard electrodes are built around a filling solution of 3 M or greater, and recommends a low ionic-strength electrode for samples below 100 µS/cm and a high ionic-strength electrode above 20,000 µS/cm. In low-conductivity water, standard electrodes drift, calibration is hard to hold and equilibration can take hours, so USGS pairs a low ionic-strength electrode with matching low ionic-strength buffers, which may not be the usual pH 4, 7 and 10 values. Check that your meter accepts them. For strongly alkaline samples, Health Canada says measurements above pH 10 "should be conducted with special 'low sodium error' electrodes."
Ohaus also sells its Starter electrodes separately for its bench and portable meters, and each includes a temperature sensor for automatic compensation. Confirm the electrode for your sample when you request a quote.
How Do You Calibrate a pH Meter?
Calibrate with at least two buffers of documented, traceable pH that bracket the pH you expect in the sample, then check the slope. That is the USGS rule, and every meter above supports it in its own way:
- Pick buffers that bracket the sample. USGS's example: if you expect pH 7 to 8, use the pH 7 and pH 10 buffers; if you expect less than 7, use pH 7 and pH 4; if the range is wide, check the electrode with a third buffer. Indicator paper is a rough guide only and is never recorded as site pH.
- Keep the buffers clean. Pour buffer into a separate container, never pour it back into the stock bottle, never dip the electrode into stock, and do not let rinse water dilute it. Bring buffers to the sample temperature. Hanna's pH calibration solutions come as 4.01, 7.01 and 10.01 in 20 mL sachets (25 per box) and in 500 mL, 1 L and 1 US gallon bottles, and the millesimal calibration solutions suit high-accuracy work.
- Check the slope. USGS accepts a slope of 95 to 102 percent. If it falls outside, clean the electrode, check the filling solution level, that the fill hole is open and that the junction is free-flowing, then recalibrate. Replace the electrode if the slope stays outside the range. For meters that show only millivolts, USGS expects −10 to +10 mV in pH 7 buffer, +165 to +195 mV in pH 4 and −165 to −195 mV in pH 10.
- Use what your meter automates. The pHep 4 and pHep 5 calibrate automatically at 1 or 2 points with two memorized buffer sets, and so do the HI991300 and HI991301. The AB23PH offers 3-point calibration with auto buffer recognition. The edge offers five-point calibration with seven pre-programmed buffers and two custom ones, and records date, time, offset, slope and the buffers used for good laboratory practice (GLP) records.
For an overview of how often instruments need professional calibration and service, see our guide to instrument calibration services.
How Do You Take a Reliable pH Reading?
Calibrate, rinse with deionized water and then with the sample, measure in still sample until the reading is stable, repeat and take the median. This sequence follows the USGS procedure for ex situ (beaker) pH measurement in surface water:
- Calibrate with bracketing buffers. Pick at least two traceable buffers either side of the expected sample pH, bring them to the sample temperature, calibrate, and check that the slope is between 95 and 102 percent.
- Rinse with deionized water. After calibration, and between different solutions, rinse the electrode and temperature sensor three times with deionized water.
- Rinse with the sample. Rinse the sensors twice with the sample: swirl one aliquot over them and discard it, then let a second aliquot sit on them for one minute.
- Measure a fresh aliquot. Pour a third aliquot into the vessel and hold the sensors in the still sample for one minute, or until the reading stabilizes within your criterion, then record it.
- Repeat and take the median. Repeat the measurement on at least two more aliquots and calculate the median of the values.
- Record pH with temperature. Record the final pH to the nearest 0.01 unit together with the sample temperature.
Health Canada adds a timing rule: measure as soon as possible after sample collection to minimize the effect of temperature.
How Do You Care for and Store a pH Electrode?
Keep liquid-filled electrodes clean, upright and wet between uses. USGS says electrodes must be clean before storage, that storage solutions are specific to the electrode type, and that glass electrodes should not be stored in deionized water unless the manufacturer says so. The bulb should be fully immersed in the storage solution, and you should not put cotton or paper towel in the cap, because it can scratch the glass membrane. Before measuring, rinse off salt deposits, unplug the fill hole, and if the electrode has been in storage solution, leave the fill hole open and suspend it in air for about 15 minutes so the junction flushes.
Hanna's electrode cleaning solutions are formulated for mineral deposits, proteins, inorganic substances and oil and fat residues on pH, ORP and ion-selective (ISE) electrodes. They come in 500 mL bottles with a 5-year unopened shelf life, and FDA-compliant bottle variants are available for food and beverage work.
Where Does pH Fit in Field and Wastewater Work?
pH is one of the routine field measurements taken with every water sample, alongside temperature, conductivity, dissolved oxygen and turbidity. For the rest of the parameter set, see our guides to portable conductivity meters, dissolved oxygen meters and turbidimeters. Conductivity in particular tells you whether a standard pH electrode is the right choice.
In treatment work pH is something you control as well as measure. If you are adjusting pH in wastewater, see our guide to pH adjustment systems for wastewater. For our full range of water-quality meters, probes and consumables, browse the Water Quality Instrumentation collection. Multi-brand supply means one quote can cover the meter, the electrode, the buffers and the cleaning solutions.
Need help choosing a pH meter?
ERE Inc. has been Canada's environmental equipment specialist for 30+ years. Tell us your sample type, accuracy target and workflow and we will match the meter, electrode, buffers and cleaning solutions.
→ Request a Quote | 1-888-287-EREC | Browse Water Quality Instrumentation | sales@ereinc.com
Frequently Asked Questions
What is a pH meter used for?
A pH meter measures how acidic or alkaline a water-based sample is. Environmental consultants use it for surface water, groundwater and wastewater checks, treatment plants use it to control chemical dosing and corrosion, and laboratories use it before other analyses. Health Canada notes that pH control matters in drinking water treatment to maximize treatment effectiveness and efficiency, and in the distribution system to control corrosion.
Which pH meter is the most accurate?
Among the instruments ERE carries, the Hanna edge lists ±0.002 pH at 0.001 pH resolution, the Ohaus Starter 2200 and Aquasearcher AB23PH list ±0.01 pH, the Hanna HI9829 lists ±0.02 pH and the Hanna pHep 5 pocket tester lists ±0.05 pH. Accuracy on the datasheet only holds with a good electrode and fresh, traceable buffers, so calibration quality decides real-world accuracy.
How do I tell if my pH meter is accurate?
Calibrate with buffers that bracket your sample pH and check the slope. USGS treats a calibration slope of 95 to 102 percent as acceptable and says to clean the electrode, check the filling solution and recalibrate if it falls outside that range. If the slope stays outside after recalibration, replace the electrode.
Does low-conductivity water need a special pH electrode?
Yes, in most cases. USGS says standard pH electrodes respond slowly in water below 100 µS/cm, with continual drift and equilibration times that can run to hours, and recommends a low ionic-strength electrode with matching low ionic-strength buffers. Check conductivity first, and confirm the electrode when you request a quote.
Related articles
- Water Quality Meters for Environmental Work: A Buyer's Guide for Canada
- Portable Conductivity Meters: How to Choose One for Field Water Monitoring in Canada
- Dissolved Oxygen Meters: How to Choose One for Field Water Monitoring in Canada
- Turbidimeter Selection Guide for Environmental Monitoring in Canada
- pH Adjustment Systems for Wastewater: Continuous-Flow vs. Batch Treatment
Lire en français : pH-mètres : comment choisir un modèle de poche, portable ou de paillasse pour l'analyse de l'eau au Canada